Transducer Kits

Accutrans Disposable Blood Pressure Transducer

Accutrans is a highly innovative and user-friendly disposable blood pressure transducer that is set to effectively meet the increasing needs of our customers.

The introduction of Accutrans adds to our wide range of medical devices in the critical care field.

Features & Benefits of Accutrans:

  • Completely compatible with current Biotrans II accessories
  • Low possibility of zero drift
  • Small, lightweight and convenient to use. Ideal for both pole and patient mount applications
  • Transparent body allows clear fluid path visibility to facilitate debubbling.
  • 2 types of flush available – controlled squeeze or pull-tab snap (dynamic waveform test)
  • Available in standard or custom kit configurations

Environmental Performance

Storage Temperature, - 30ºC to + 70ºC
Operating Temperature, + 15ºC to + 45ºC
Operating Humidity, 10% to 90% non condensing
Operating Atmospheric Pressure, 400 mmHg to 850mmHg

Mechanical Specifications

Pressure Range, -30 to 300 mmHg
Overpressure Withstand, 125 PSI
Mounting, Any axis

Electrical Performance

Transducer Excitation Voltage, 1 to 10 Vrms
Transducer Signal Frequency, DC 5000Hz
Phase Shift, < 5 degree C
Transducer Excitation Impedance, 1200 to 2000ohms
Transducer Signal Impedance, 285 to 375 ohms
Transducer Symmetry, ± 5%
Accuracy, Meets AAMI/ANSI BP22 (1994) standard
Sensitivity, 4.95-5.05 μ 5V/V/mmHg Nomina
Zero Drift after 5 min warm-up, < 1 mmHg in 4 hrs
Temperature Sensitivity Coefficient, < 0.1% / degree C
Zero Offset, ± 20mmHg
Light Sensitivity, < 1 mmHg
Mechanical Shock (2ms sine, 3 axis), 500 g
Vibration (20Hz to 2kHz, 1 ocrave/min, 3 axis), 20 g

Safety Features

Risk Current, Less than 5mA
Defibrillation Withstand, 5 cycles in 5 minutes at 360J

Catalog/ Reorder Number & Description

HM650-150, Vertical Mounting Clamp for mounting up to 3 transducers holders.
HM650-150-H, Horizontal Mounting Clamp for IV poles up to 1-1/2 inches (4cm) diameter. Holds up to 3 transducer holders.
MAC-XX, 10 foot (3.05m) Monitor Adapter Cables for specific brands of monitors. Specify monitor brand (and amplifier plug-in), model numbers, and provide a drawing of the pins and connector shape at the pressure input on the monitor. Contact your Biosensors representative for assistance. See ACCUTRANS Accessory catalog for details.
PMA-1, Replacement Pole-Mount Accessory Docking Bracket/Transducer Holder